Posts Tagged ‘pressure recovery phenomenon’

Bernoulli principle states  that , when a high pressure jet (Air, Water, blood etc ) moves over a conduit,  the pressure exerted by the jet on its sides (Lateral wall) reduces . The velocity gain is equal to pressure drop .This is why we take velocity as a rough guide to pressure gradient and the sacred formula in doppler echocardiography 4V2 came in to vogue . (Incidentally, Bernoulli principle shares the same principle when aircrafts  lifts from runway at its peak speed as the pressure above the wings   drops to zero or negative  and the plane lifts up.)

Please note , the pressure should drop both above and below the aircraft by Bernoulli principle .But, the engine and wings are arranged in such a way , the air speed below the aircraft is slower and hence the pressure is high below and low above and the lift occurs promptly at take of velocity. Imagine , how the valve leaflets in heart is subjected to lift and drag forces every time the blood gushes with high velocity flows.This is also the reason for the Pulsus bisferiens, SAM in HOCM, Coanda effect in supra valvular stenosis, and any post stenotic dilatation.

 

In Echocardiography the Bernoulli equation is modified.

In clinical doppler echocardiography, we have liberally simplified the original Bernoulli equation by ignoring the the proximal sub valvular velocity V1 . Further , two more components in the equation is also amputated  for our convenience ! (Flow acceleration and the viscous friction) .This is the reason we tend to err many times  especially in outflow tract gradients and prosthetic valve gradients .

Pressure recovery phenomenon.

This  is another  hemodynamic lacunae in clinical echocardiography. We know, thepeak velocity of blood is attained  just distal to site of  obstruction. As the distal velocity beyond the obstruction begins to fall, the pressure tends to recover corresponding to the loss of velocity. This happens to certain distance beyond the obstruction. Since continuous wave doppler measures the pressure in its entire axis of alignment , it  is likely to pick more pressure samples  from the recovered areas and net result is, it measures more than the true difference in gradient across the valve.The phenomenon is most relevant in assessment of Aortic stenosis and results in over estimation of severity of stenosis.

How much can be the overestimation ?

It can be up to 30 % or even more.Especially in prosthetic Aortic valves.

How to recognise it and overcome it ?

  1. First of all,  recognise such a hemodynamic phenomenon  exists and the sacred 4v square can be a myth !
  2. Never go with gradient alone in diagnosing valve stenosis. Look for 2D features also.This is more vital when you suspect acute valve obstruction.
  3. Always add the proximal sub valvular velocity (V1 ) in your Bernoulli equation .It need to be subtracted.
  4. The effect of heart rate on pressure recovery has not been properly studied.(The impact of which  could be vital and hence too many false prosthetic emergencies could be avoided, as cardiologists tend to rely mostly on gradient than anatomical diagnosis of valve obstruction like visualising thrombus or struck leaflet by TEE or fluro.

Does this phenomenon happen with cath gradient ?(Generally it’s more pronounced in doppler echo )

Yes, It does happen in cath lab also , as its related to physics of flow. It can be minimised if we can use two simultaneous catheters ,one in LV  and the other Aortic catheter placed very close to the leaflets.

pressure recovery in aortic stenosis animation

Click below for an Animated version

pressure recovery phenomenon in aortic stenosis 005

Note the pressure recovers from P 2 to P3

Reference

Pressure recovery phenomenon in doppler echocardiography

pressure recovery phenomenon doppler echocardiography